AWWA Water Science Фильтрация и сепарация
Method Development, Interlaboratory Study, and Quality Control Criteria for Flow Imaging Microscopy Analysis of Phytoplankton in Freshwater
Аннотация
ABSTRACT Flow imaging microscopy (FIM) is used for rapid characterization of phytoplankton in water; however, standardized quality control (QC) and interlaboratory validation are lacking. This study presents the development of Standard Method 10210 for FIM: method optimization, QC criteria, and a multi‐laboratory validation study (10 laboratories and 13 analysts). Standardized protocols were established using microbead reference materials and applied to source water samples. Interlaboratory testing demonstrated strong reproducibility, with relative standard deviations (RSDs) for non‐fluorescent microbeads analyzed in AutoImage Mode < 10% (30–70 μm) and < 20% (10 μm). Fluorescent microbead RSDs were < 20% (AutoImage Mode) and < 25% (Trigger Mode). Surface water sample RSDs were ≤ 30% in AutoImage Mode across triplicates, days, and laboratories. The study established QC capable of identifying operational issues and demonstrated consistent classification of phytoplankton across laboratories/holding times. This work provides the first validated QC framework for FIM‐based monitoring, enabling reliable, rapid, and standardized assessment of phytoplankton populations for source management and early bloom detection.